A super‐resolution proof of concept in a cycloidal coded aperture miniature mass spectrometer

Rationale

Higher resolution in fieldable mass spectrometers (MS) is desirable in space flight applications to enable resolving isobaric interferences at m/z < 60 u. Resolution in portable cycloidal MS coupled with array detectors could be improved by reducing the slit width and/or by reducing the width of the detector pixels. However, these solutions are expensive and can result in reduced sensitivity. In this paper, we demonstrate high-resolution spectral reconstruction in a cycloidal coded aperture miniature mass spectrometer (C-CAMMS) without changing the slit or detector pixel sizes using a class of signal processing techniques called super-resolution (SR).

Methods

We developed an SR reconstruction algorithm using a sampling SR approach whereby a set of spatially shifted low-resolution measurements are reconstructed into a higher-resolution spectrum. This algorithm was applied to experimental data collected using the C-CAMMS prototype. It was then applied to synthetic data with additive noise, system response variation, and spatial shift nonuniformity to investigate the source of reconstruction artifacts in the experimental data.

Results

Experimental results using two ½ pixel shifted spectra resulted in a resolution of ¾ pixel full width at half maximum (FWHM) at m/z = 28 u. This resolution is equivalent to 0.013 u, six times better than the resolution previously published at m/z = 28 for N2+ using C-CAMMS. However, the reconstructed spectra exhibited some artifacts. The results of the synthetic data study indicate that the artifacts are most likely caused by the system response variation.

Conclusions

This paper demonstrates super-resolution spectral reconstruction in C-CAMMS without changing the slit or detector pixel sizes using a sampling SR approach. With improvements, this technique could be used to resolve isobaric interferences in a portable cycloidal MS for space flight applications.

Tanouir Aloui,
Rafael Bento Serpa,
Nabil Abboud,
Kathleen L. Horvath,
Justin Keogh,
Charles B. Parker,
Jennifer C. Stern,
M. Bonner Denton,
Maria Luisa Sartorelli,
Jeffrey T. Glass,
Michael E. Gehm,
Jason J. Amsden
February 10, 2023
https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/10.1002/rcm.9477?af=R