Linked scan at constant B/E

From Mass Spec Terms
(Redirected from B/E Linked Scan)
DRAFT DEFINITION
Linked scan at constant B/E
Related Term(s): Linked scan
Reference(s):

IUPAC. Analytical Division. Compendium of Analytical Nomenclature (the Orange Book). Definitive Rules, 1979. Compiled by J. Inczédy, T. Lengyel, A. M. Ure. Blackwell Scientific Publications, Oxford (1997). On-line corrected version: http://www.iupac.org /publications/analytical compendium (2000).

This is an unofficial draft definition presented for information and comment.

Recommended terms | Full list of terms

 

Orange Book

ORANGE BOOK DEFINITION

IUPAC. Analytical Division. Compendium of Analytical Nomenclature (the Orange Book). Definitive Rules, 1979 (see also Orange Book 2023)

Linked scan at constant B/E

This can be performed with a sector instrument incorporating at least one magnetic sector and one electric sector. The magnetic sector field strength, B, and the electric sector field strength, E, are scanned simultaneously, while holding the accelerating voltage, V, constant, so as to maintain the ratio B/E constant. This constant value is determined by the ratio of the two field strengths required to transmit main-beam ions of predetermined mass/charge ratio. These preselected main-beam ions are the precursor ions whose fragment ion spectrum is required. The term B/E linked scan is not recommended.

IUPAC 1997 Orange Book Chapter 12